var fDesc=new Array(); fDesc[0] = "Ansur plug-in module to control the QA-90 Mk II Electrical Safety Analyzer.
The QA-90 Test Sequence has a parameter called "Power off delay time" which can be used to add a delay at the end of the sequence. When using this feature, you will need to add 6 seconds to any delay you specify. This is because the QA-90 incorrectly processes the value being provided for this delay. For example, if a 2 second delay is desired, use 8 seconds instead. If 5 seconds of delay is required, enter 11 as the value. Using a delay of 6 or less will result in no delay happening at all."; function tShowHide(id, show) { var s = document.getElementById("desc"); if ((s.innerHTML.length<=212 || show==1) && show!=2) { s.innerHTML = fDesc[id]; if (document.getElementById('m1')) document.getElementById('m1').style.display='none'; if (document.getElementById('m2')) document.getElementById('m2').style.display='none'; if (document.getElementById('more_txt')) document.getElementById('more_txt').style.display='inline'; } else { s.innerHTML = ''; } }